CF measuring amplifier MBI 50.25.
1-channel carrier frequency measuring amplifier for operation of inductive transducers / LVDT sensors in die-cast field housing
- Proven analogue circuit for industry and laboratory
- Robust design on eurocard PCB or in field housing
- System-compatible for 19"-housing and 19" sub-racks
Important features
- Supply: 230 VAC
- Carrier frequency: 10 kHz
- Output signal: ±10 V or 4...20mA
- Design: Die-cast field housing
Suitable sensors
Type | Inductive transducers in differential LVIT sensor or differential transformer circuits | |
Required sensitivity | 10 ... 400 mV/V |
Oscillator
Carrier frequency |
approx. 10 kHz (trapezoidal)
|
|||||
Bridge supply voltage |
ca. 1 Vrms
|
Amplifier
Linearity error |
< 0.1%
|
|||||
Temperature coefficient of zero point | < 0.1 % / 10 K | |||||
Temperature coefficient of sensitivity
|
< 0.2 % / 10K | |||||
Interference voltage (residual carrier voltage)
|
< 2 mVrms
|
|||||
Load capacity max.
|
10 mA
|
|||||
Drift
|
ca. ±10 % of final value; up to ±100 % possible as special version
|
|||||
Cutoff frequency for the measurement signal (-3 dB)
|
500 Hz
|
General information
Max. current | 25 mA at ±15 V supply without current output stage 50 mA at ±15 V supply with current output stage |
|||||
Zero-point adjustment | through potentiometer | |||||
Gain adjustment | fine: by potentiometer coarse: by fixed resistor |
|||||
Phase adjustment | through fixed capacitor | |||||
Electrical connector | Plug connector according to DIN 41612, 32-pin, type C Special version: 16-pin terminal block |
|||||
Required mating connector | Female connector according to DIN 41612, 32-pin, type C or D | |||||
Operating temperature | 0 ... 60 °C | |||||
Storage temperature | -25 ... 85 °C |
Type-dependent information
Operating voltage | ±15 V; 230 V~ im Feldgehäuse | |||||
Rated output | ±10 V; 4 ... 20 mA | |||||
Dimensions without front panel and field housing | ca. 100 x 167 x 25 mm | |||||
Front panel dimensions | 35,3 x 128,4 mm (19": 7 TE x 3 HE) | |||||
Weight without front panel and field housing | ca. 0,2 kg | |||||
Weight with front panel | ca. 0,23 kg |
Version overview
Product code | Casing | Supply | Output signal |
MBI 50.25.10
|
Die-cast field housing
|
230 V~
|
10 V
|
MBI 50.25.12
|
Die-cast field housing
|
230 V~
|
4 ... 20 mA
|
Note: The 10 V output provides a Nominal output signal of ±10 V for balanced systems (these are the most common inductive transducers). With unbalanced systems, the output is 0 ... 10 V.
Do you have further technical questions about the product?
We will be happy to help you determine the order code and provide you with further specific information for your required product.